Nippon Medical & Chemical Instruments Co., Ltd. is a manufacturer specializing in environmental control, medical device, and scientific instruments.
Wide Wave LED (WLED)
Six individual rooms can be separately controlled for temperature and illumination.
Wide Wave LED (WLED) are used to save the space, energy, and cost, as well as to speed up the research & development.
Model | LH-80WLED-6CT |
---|---|
Outer dimensions (WxDxH) [mm] | 1,555×830×1,985 |
Inner dimensions (WxDxH) [mm] | 395×505×281×6 rooms |
Control system | Changeover system between constant operation and day/night switching operation |
Material | Outer panel:stainless steel coated with resin, inner panel:stainless steel |
Temperature | +5~45℃ During lighting:+15~40℃, temperature difference not more than 20℃ |
Light source | Wide Wave LED (WLED); RA90, the number of LED chips:126, dimmer control |
Illumination | 0 to 8,800 lx (at the shelf net) |
Photon amount | 0 to 135 μmolm-2s-1(at the shelf net) |
Power requirement | Single phase 100V 18A/Single phase 100V 15A (for a power-outlet installed in the incubator) |
Weight | About 250kg |
# <Temperature setting of LH-80WLED-6CT>
# It is advisable to install the Biomulti incubator in a commonly-used room below 30℃.
# Temperature may vary during defrosting.
# There are some limitations in temperature settings. See the instruction manual.
# Temperature difference between the highest and the lowest should be set to be not higher than 20℃. If higher than 20℃, the temperature may be uncontrollable. In this case, some measures should be taken, such as reducing the number of rooms to be controlled.
# Sum of the differences between the set temperatures and lowest temperatures should be less than 60-80. If the lowest setting temperature is larger, this value tends to be larger. If the temperature is set to be low or the light is turned on, this value should be lower.
Example)A
(30-10)+(30-10)+(20-10)+(20-10)+(10-10)+(10-10)=60≦(60~80)
A | |
30℃ | 30℃ |
20℃ | 20℃ |
10℃ | 10℃ |