Foldable Negative Pressure Booth for Infection Control

For prevention of nosocomial infection, new influenza infection, etc.

Features

  • Easy to be installed with simple setting of ducts (* can also be used without ducts).
  • Keeps negative pressure inside the chamber, and does not leak contaminated air into environment.
  • Quickly sucks contaminated air with wide suction areas.
  • HEPA filters are used in the exhausting system.
  • Anti-bacterial painting on the body.
  • Transparent plastic curtain without oppressive feeling (flame-proof, antistatic treatment).
  • Quiet operation.
  • A reading light can be equipped (option).
  • Can be folded and compactly stored when not used.
  • You can watch the installation movie here (You tube)

Options

  • Overhead reading light

Specifications

Model NB-1200S
Outer dimensions (W × D × H) (㎜) 1,274 × 2,635 × 1,872
Inner dimensions (W × D × H) (㎜) 1,160 × 2,284 × 1,841
Filter  Dust collection efficiency of HEPA filter: PAO particle 0.3μm, 99.99% or more
Air volume flow  rate 9 or 4.5m3/min, two step changing-over
Ventilation frequency  100 or 50 times/h
Internal pressure  Negative
Noise 40 dB(A) at the lower air volume flow rate  (for reference only)
Power requirement  Single Phase 100 V 50/60 Hz
Power consumption 200 W

* Attached with a video explaining the installation.